Article Dans Une Revue Clinica Chimica Acta Année : 2022

Performance evaluation of UF-4000 body fluid mode for automated body fluid cell counting

Résumé

Background: Cytological analysis of body fluids (BF) provides important information for diagnosis in various medical conditions. We evaluated the analytical performance of the UF-4000 BF mode for ascitic, cerebrospinal, pleural, synovial and continuous ambulatory peritoneal dialysis fluids compared to light microscopy counting (LM). Materials and methods: 223 consecutive BF were analyzed by UF-4000 and results were compared using Pearson's correlation, Bland-Altman analysis, and contingence tests at relevant cut-off values. This study also included the evaluation of precision, linearity, and carryover. Results: For white and red blood cells (WBC, RBC) counts in all BF, correlation was excellent with Pearson's coefficients R2 > 0,98. Bland-Altman analysis didn't reveal significant differences with limited bias for WBC ranging from -10 to -1 WBC/µL and bias ranging from -43 to -6/µL for RBC. At specific cut-off values for WBC, Se and Spe were 100% except for ascites (Spe = 98%) due to two false positive. Precision evaluated at three concentration levels was good for each parameter (WBC < 10%). Linearity was excellent for WBC (R2 > 0,99) and carryover negligible (<0,004%). Conclusion: UF-4000 BF mode is a good alternative to manual LM for BF cell counting. This automated method gives rapid and accurate results which is important for therapeutic decisions.
Fichier principal
Vignette du fichier
1-s2.0-S000989812200122X-main.pdf (353.16 Ko) Télécharger le fichier
Origine Fichiers produits par l'(les) auteur(s)
Licence

Dates et versions

hal-03706354 , version 1 (24-05-2024)

Licence

Identifiants

Citer

Marie Siatkowski, Sandrine Dahyot, Martine Pestel-Caron, Sophie Boyer. Performance evaluation of UF-4000 body fluid mode for automated body fluid cell counting. Clinica Chimica Acta, 2022, 531, pp.152-156. ⟨10.1016/j.cca.2022.04.002⟩. ⟨hal-03706354⟩
47 Consultations
2 Téléchargements

Altmetric

Partager

Gmail Mastodon Facebook X LinkedIn More