Pulse shaping optimization for improving atom probe tomography - Normandie Université
Article Dans Une Revue International Journal of Mass Spectrometry Année : 2015

Pulse shaping optimization for improving atom probe tomography

Résumé

We propose in this paper novel approach for improving mass resolution in atom probe tomography. Using conventional counter-electrode or microelectrode design, improved mass resolution at half, tenth and 1% of the mass peak maximum is predicted when shaping properly the voltage evaporation pulse. Using a numerical approach, it is shown that a flat top voltage pulse used to trigger the field evaporation with sharp front and leading edges associated with a short tip to counter electrode distance(≈10 μm) strongly minimizes the energy deficits of evaporated ions from the sample, so that energy compensation devices are not necessary to obtain high mass resolution.
Fichier non déposé

Dates et versions

hal-02177749 , version 1 (09-07-2019)

Identifiants

Citer

Lu Zhao, Antoine Normand, Fabien Delaroche, B. Ravelo, François Vurpillot. Pulse shaping optimization for improving atom probe tomography. International Journal of Mass Spectrometry, 2015, 386, pp.47-53. ⟨10.1016/j.ijms.2015.06.012⟩. ⟨hal-02177749⟩
26 Consultations
0 Téléchargements

Altmetric

Partager

More