Résumé

This study is concerned with the design of a Mueller imaging polarimeter for the visualization of spatially-varying Mueller matrix fields. A simplified calibration procedure is advocated, where all the optical elements are calibrated simultaneously rather than independently as in the state-of-the-art. This is shown to significantly reduce the bias inherent to sequential calibration methods. In addition, this procedure requires no reference sample, it allows calibration both in transmission or in reflection modes, and it relies on ready-to-use cameras. Put together, these novelties should help non-specialists in optics designing and calibrating a Mueller imaging polarimeter for applications such as material classification.

Fichier principal
Vignette du fichier
paper.pdf (6.9 Mo) Télécharger le fichier
Origine Fichiers éditeurs autorisés sur une archive ouverte
Licence
Loading...

Dates et versions

hal-02169371 , version 1 (01-07-2019)

Licence

Identifiants

Citer

Yvain Quéau, Florian Leporcq, Ayman Alfalou. Design and simplified calibration of a Mueller imaging polarimeter for material classification. Optics Letters, 2018, 43 (20), pp.4941. ⟨10.1364/OL.43.004941⟩. ⟨hal-02169371⟩